Able to Withstand against 10 Times Larger Overload than Rated Capacity by Built-in Overload Stopper
AS-TG series acceleration transducers have 3 incorporated miniature acceleration transducers to measure simultaneously 3 axial accelerations (front-rear, left-right, and up-down directions). Not only are interferences of AS-TG series among each axes little, but also AS-TG series can withstand against 10 times larger overload than rated capacity.
Moreover, it can be used also to 10,000,000 times of repetition loads.
Model |
Rated Capacity
(Reference Value) |
Frequency Response
(at 23??C ) |
AS-1TG |
??9.807m/s 2 (??1G) |
DC to 40Hz ??5% |
AS-2TG |
??19.61m/s 2 (??1G) |
DC to 60Hz ??5% |
AS-5TG |
??49.03m/s 2 (??1G) |
DC to 100Hz ??5% |
AS-10TG |
??98.07m/s 2 (??1G) |
DC to 150Hz ??5% |
AS-20TG |
??196.1m/s 2 (??1G) |
DC to 250Hz ??5% |
Performance |
Rated Capacity |
See table above |
Nonlinearity |
Within??0.1% RO |
Hysteresis |
Within??0.1% RO |
Rated Output |
0.5 mV/V (1000??m/m) or more |
Environmental Characteristics |
Safe Temperature Range |
-10 to 60??C |
Electrical Characteristics |
Safe Excitation Voltage |
6V AC or DC |
Recommended Excitation Voltage |
1 to 3V AC or DC |
Input Resistance |
121???1.7% |
Output Resistance |
121???1.7% |
Cable |
4-conductor (0.08mm 2 ) vinyl shielded cable diameter 3.2 mm by 5 m long, terminated with connector plugs |
(Shield wire is not connected to mainframe.) |
Mechanical PropertiesConnector |
Safe Overload Rate |
1000% (with stopper) |
Frequency Response Range |
See table above |
Transverse Sensitivity |
??4% |
Weight |
Approx. 110 g (Excluding cable) |
Note:
Measurement range of up-down direction is narrowed by 1G (9.807m/s 2 )
because gravity acceleration is always applied to acceleration
transducers. In terms of AS-1TG, detected value exceeds rated capacity
in positive side toward gravidy direction. Although AS-TG
acceleration transducers are not broked under safe overload rating,
characteristics are out of warranty.