    Hermetically-seal Structure with Inert Gas Filled in Usable for Both Tension and Compression Loads

    The detection portion is hermetically sealed with inert gas filled in to prevent aging deterioration and to ensure reliability and stability for a long period of time.
    Features
    * TEDS-installed versions can be manufactured. Inquiries are welcome.
    PDF Download
    Specifications
    Model Rated Capacity Natural Frequency
    (Approx.)
    Weight
    (Approx.)
    Saddle Mount Base
    LU-50KE ??500N 1.54kHz 2.8kg CA-1B CF-50
    LU-100KE ??1kN 2.16kHz
    LU-200KE ??2kN 3.28kHz
    LU-500KE ??5kN 2.66kHz 2.8kg CA-1B CF-80
    LU-1TE ??10kN 4.2kHz 2.8kg N/A N/A
    LU-2TE ??20kN 4.97kHz 2.8kg
    LU-5TE ??50kN 3.5kHz 5.0kg
    LU-10TE ??100kN 3.14kHz 9.5kg
    LU-20TE ??200kN 2.5kHz 22.0kg
    Performance
    Rated Capacity See table above
    Nonlinearity Within??0.2% RO
    Hysteresis Within??0.1% RO
    Repeatability 0.1% RO or less
    Rated Output 2 mV/V (4000??m/m) ??0.2%
    Environmental Characteristics
    Safe Temperature Range -30 to 85??C
    Compensated Temperature Range -10 to 70??C
    Temperature Effect on Zero Balance Within??0.005% RO/??C
    Temperature Effect on Output Within??0.005%/??C
    Electrical Characteristics
    Safe Excitation Voltage 20V AC or DC
    Recommended Excitation Voltage 1 to 10V AC or DC
    Input Resistance 350???0.5%
    Output Resistance 350???0.5%
    Cable 4-conductor (0.3 mm 2 ) chloroprene shielded cable,
    7.6 mm diameter by 5 m long, terminated with NDIS connector plug
    (Shield wire is not connected to mainframe.)
    Mechanical Properties
    Safe Overload Rating 150%
    Natural Frequency See table above