| Model |
Rated Capacity |
Natural Frequency
(Approx.) |
| PGR-10KA |
1MPa |
12kHz |
| PGR-20KA |
2MPa /td>
| 17kHz |
| PGR-50KA |
5MPa |
29kHz |
| PGR-100KA |
10MPa |
42kHz |
| PGR-200KA |
20MPa |
60kHz |
| Performance |
| Rated Capacity |
See table above |
| Nonlinearity |
Within??0.1% RO |
| Hysteresis |
Within??0.1% RO |
| Rated Output |
1.5 mV/V (3000??m/m) ??5% |
| Environmental Characteristics |
| Safe Temperature Range |
-30 to 110??C |
| Compensated Temperature Range |
-10 to 100??C |
| Temperature Effect on Zero Balance |
Within??0.01% RO/??C |
| Temperature Effect on Output |
Within??0.01%/??C |
| Electrical Characteristics |
| Safe Excitation Voltage |
12V AC or DC |
| Recommended Excitation Voltage |
1 to 8V AC or DC |
| Input Resistance |
350???1.4% |
| Output Resistance |
350???1.4% |
| Cable |
4-conductor (0.75 mm 2 ) fluonlex shielded cable,
8 mm diameter by 5 m long, bared at the tip
|
| (Shield wire is not connected to mainframe.) |
| Mechanical PropertiesConnector |
| Safe Overload Rating(*1) |
300% |
| Critical Overload(*2) |
117.7 MPa (PGR-10 to 50KA)
196.1 MPa (PGR-100 & 200KA)
|
| Natural Frequency |
See table above |
| Weight |
Approx. 400 g |
| Mounting Screw |
G3/8, male |
* 1. Maximum overload which can be applied without causing any permanent change in
specified characteristics
* 2. Maximum overload which can be applied without causing any structural damage.