    Heat-resistant Sputter Gages Achieve Pressure Measurement at High-temperature

    To enable pressure measurement at high temperature, PHC-B series pressure transducers adopt thin-film strain gage formed by sputtering.The sensor part is a flush, diaphragm detecting pressure directly on a flat surface without pressure medium, thus enabling pressure measurement without missing momentary pressure changes. In addition, the flush diaphragm makes these transducers suitable for measuring not only liquid or gas pressure but also pressure of highly viscous medium.The small-sized design and flexible cable make them easy to use even in limited space.
    Features
    ?? Safe temperature range from -30 to 240??C
    ?? Flush diaphram ensuring high frequency response
    ?? Compact, flexible, and heat-resistant cable ensuring ease of use
    PDF Download
    Application
    Model Rated Capacity Natural Frequency
    (Approx.)
    PHC-B-2MP 2MPa 45kHz
    PHC-B-5MP 5MPa 75kHz
    PHC-B-10MP 10MPa 85kHz
    PHC-B-20MP 20MPa 85kHz
    Performance
    Rated Capacity See table above
    Nonlinearity Within??0.5% RO
    Hysteresis Within??0.3% RO
    Repeatability 0.2% RO or less
    Rated Output 0.6 mV/V (1200??m/m) or more
    Environmental Characteristics
    Safe Temperature Range -30 to 240??C (200??C with cable,-25 to 80??C with connector plug)
    Compensated Temperature Range 23 to 230??C
    Temperature Effect on Zero Balance Within??0.03% RO/??C
    Temperature Effect on Output Within??0.03%/??C
    Electrical Characteristics
    Safe Excitation Voltage 12V AC or DC
    Recommended Excitation Voltage 1 to 10V AC or DC
    Input Resistance 380 to 650??
    Output Resistance 380 to 650??
    Cable 4-conductor (0.09 mm 2 ) fluoroplastic shielded cable,
    3.1 mm diameter by 3 m long, terminated with connector plug
    (Shield wire is not connected to mainframe.)
    Mechanical PropertiesConnector
    Safe Overload Rating 150%
    Natural Frequency See table above
    Material Case: SUS metallic finish
    Liquid-contacting part: SUS 630
    Weight Approx. 70 g (not including cable)
    Mounting Screw G1/8, male
    Standard Accessories
    Gasket (mild copper)

    Back to top