LAT-1000A Series
Enables Highly Accurate Measurement Possible to Inference Correction by Arithmetic Processing.
Each system in the LAT-1000A series consists of the LAT-A 6-component force transducer and the FDP-106A signal
processor.
The LAT-A simultaneously detects 3 forces in 3 axial directions orthogonal to the transducer and 3 moments around the 3 axes. The FDP-106A automatically eliminates interference components contained in transducer output through calculation.
By minimizing errors due to interference, the system enables highly accurate measurement of both single and multiple component force loads. (Patented)
?? 5-V output available with the rated load
?? To guarantee measurement accuracy, performance
with multiple component force loaded is indicated with a maximum error (see note in the next page).
?? Highly accurate measurement possible even with multiple component force loaded
?? Simultaneous sampling of 6-component force and processing signals
up to approximately 300 Hz possible
?? The compact, lightweight transducer is strain gage based and is cased
with a highly rigid special aluminum alloy.
?? Calibration coefficient is preset in the signal processor, enabling immediate measurement
by connecting a monitor indicator.
?? Force and moment can be read directly on a PC if connected.
?? Direct reading mode is provided to read force and moment at the load point.
?? High/low limit and hysteresis width of the high/low limit are set to alarm output.
?? 6-component force transducer LAT-A
?? Signal processor FDP-106A
?? PC (not included)
Rated Capacity |
|
Model |
Rated Capacity |
Fx
N |
Fy
N |
Fz
N |
Mx
N・m |
My
N・m |
Mz
N・m |
LAT-1030KA-1 |
300 |
300 |
300 |
10 |
10 |
10 |
LAT-1030KA-2 |
300 |
300 |
300 |
20 |
20 |
20 |
Safe Overload Rating |
120% |
Nonlinearity |
Within??0.5% RO |
Hysteresis |
Within??0.5% RO |
Interference |
??0.8% RO |
Maximum Error |
??1.5% RO (??3% RO with LAT-KA-2) |
Resolution |
0.05% FS |
Temperature Effect on Zero Balance |
Within ??0.25% RO/??C |
Temperature Effect on Output |
Within ??0.05%/??C |
Compensated Temperature Range |
0 to 50 ??C |
Specifications stated above are values measured with our calibrators under in company standard conditions.
6-Component Force Transducer LAT-A Specifications
Rated Capacity |
Fx, Fy=300N, Fz=300N |
Mx, My, Mz=10, 20N・m |
See table in the previous page for combinations. |
Safe Overload Rating |
120% |
Natural Frequency (with all models) |
Fx??Fy : Approx. 2.3 kHz, Fz: Approx. 5.5 kHz |
Mx??My : Approx. 8 kHz, Mz: Approx. 4 kHz |
Recommended Excitation Voltage |
2.5V??DC |
Safe Excitation Voltage |
5V?? DC |
Input Resistance |
58.3?????10% |
Output Resistance |
350??? ??2% |
Compensated Temperature Range |
0 to 60??C |
Safe Temperature Range |
0 to 70??C |
Temperature Effect on Zero Balance |
Within 0.05% RO/??C |
Temperature Effect on Output |
Within 0.05%/??C |
Weight |
Each model approx. 250 g (not including cable) |
Protection Rating |
IP30 (JIS C 0920) |
Cable |
14-conductor (0.3 mm2 ) PVC shielded cable, 9 mm diameter,with connector plug at both ends. |
N-78 for connection to FDP-106A (Shield wire is not connected to mainframe) |
For measurement of displacement and inclinaton angle, contact us . |
Standard Accessories |
Communications program (Windows version), torque wrench, hexagon socket wrench, parallel pins ??4 and ??8, connection cable N-78 |