  • LSM-B-SA1

    Enables Force Measurement in X, Y and Z Directions. The Compact, Lightweight Strain Gage Based Design is Suitable for Model Experiments. (Patented)

    10 to 500N
    Features
    Notes : 1. Mx, My and Mz cannot be measured.
                2. Arrows indicate directions of component force in plus 
                    polarity acting to the B plane with the A plane fixed.
    PDF Download
    Specifications
    Model Rated Capacity Natural Frequency
    (Approx.)
    Safe Moment
    (Approx.)
    Weight(Approx.)
    Fx, Fy, Fz X,Y Z Mx,My,Mz
    LSM-B-10NSA1 10N 0.3kHz 0.2kHz 1.2N??m 600g
    LSM-B-20NSA1 20N 0.4kHz 0.3kHz 2.4N??m
    LSM-B-50NSA1 50N 0.8kHz 0.6kHz 5.9N??m
    LSM-B-100NSA1 100N 1.3kHz 0.9kHz 9.8N??m
    LSM-B-200NSA1 200N 2.5kHz 2.0kHz 24N??m
    LSM-B-500NSA1 500N 2.2kHz 1.8kHz 59N??m 1.6kg
    Performance
    Rated Capacity See table above
    Nonlinearity Within ??0.5% RO
    Hysteresis Within ??0.5% RO
    Rated Output Approx. 0.5 mV/V (1000??m/m)
    Environmental Characteristics
    Safe Temperature Range 0 to 80??C
    Compensated Temperature Range 0 to 70??C
    Temperature Effect on Zero Balance Within??0.05% RO/??C
    Temperature Effect on Output Within??0.05%/??C
    Within??0.05%/??C
    Safe Excitation Voltage 10V AC or DC
    Recommended Excitation Voltage 1 to 5V AC or DC
    Input Resistance 240???5%
    Output Resistance 240???5%
    Cable 4-conductor (0.3 mm 2 ) chloroprene shielded cable,
    4 mm diameter by 5 m long, with connector plug to
    the transducer side and bared to the amplifier side
    (Shield wire is not connected to mainframe.)
    Mechanical Properties
    Mechanical Properties 150%
    Natural Frequency See table above
    Weight See table above
    Interference ??3%RO
    Safe Moment See table above
    * For the cable terminated with NDIS connector, suffix "-P" to the model number.