LCH-F
Can Measure Compression Loads with 1/5000 Accuracy.
Useable in high humidity.
?? Remote sensing possible
?? Watertight structure
Model |
Rated Capacity |
Natural Frequency
(Approx.) |
Weight
(Approx.) |
Movable
Saddle |
LCH-10TF |
100kN |
7.5kHz |
12kg |
ER-10B |
LCH-20TF |
200kN |
7kHz |
17kg |
--- |
Performance |
Rated Capacity |
See table above |
Nonlinearity |
Within??0.02% RO |
Hysteresis |
Within??0.02% RO |
Repeatability |
0.02% RO or less |
Rated Output |
2 mV/V (4000??m/m) ??0.1% |
Environmental Characteristics |
Safe Temperature Range |
-35 to 80??C |
Compensated Temperature Range |
-10 to 60??C |
Temperature Effect on Zero Balance |
Within ??0.0015% RO/??C |
Temperature Effect on Output |
Within ??0.001%/??C |
Electrical Characteristics |
Safe Excitation Voltage |
20 VAC or DC |
Recommended Excitation Voltage |
1 to 10 VAC or DC |
Input Resistance |
350???0.5% |
Output Resistance |
350???0.5% |
Cable |
6-conductor (0.5 mm 2 ) chloroprene shielded cable,
9.5 mm diameter by 5 m long, with press-fit terminal for 4 mm |
(Shield wire is not connected to mainframe.) |
Mechanical Properties |
Safe Overload Rating |
200% |
Natural Frequency |
See table above |
Weight |
See table above |
Protection Rating |
IP67 (watertight type conforming to JIS C 0920) |